Change an ac-9 Plus from Data Logger to ac-9 Mode

  1. Start WL Host v. 7.09 and appropriate PC COM port at the Logger Port Setup area (highlighted).
  2. Supply power to meter.

Within several seconds the host program (right) displays:

  • COM Port
  • Baud rate
  • Logger Status—The meter's current flash setting is displayed (Moored Logging)

Select Advanced User from the Modes menu at the top of the window.
Select Get Setup. The resulting Port Setup and Logger Control tabs are shown below.


Port Setup tab, Moored Logging mode

To switch from Data Logger mode to Direct To ac9 without cycling power:
1. Go to Logger Control tab, select Stop.
2. Select Get Setup.
3. Change Mode of Operation (right) to Run Logger Port.
4. Select Send Setup. The resulting Port Setup and Logger Control tabs are shown below.


Port Setup tab in Direct To ac9 mode


Logger Control tab in Direct To ac9 mode.

Change ac-9 Plus from ac-9 to Logger mode