ac-9 Plus meters can operate as either a straight ac-9 or as a data logger (DH-4) depending on the meter's flash memory setup. Meters are typically sent from the factory in Data Logger mode unless a different setup is requested.
Several seconds later, the host program (right) displays:
|
![]() |
3. Select Advanced User mode from the Modes menu at the top of the window. Select Get Setup . View the settings retrieved from the meter on the Port Setup and Logger Control tabs (shown below).

Port Setup tab, Direct To ac9 mode

Logger Control tab, Direct to ac9 mode
To change from Direct To ac9 to Data Logger mode without cycling power to the meter, select the Logger Control tab, then the desired Mode of Operation. Select Send Setup.

Logger Control tab, Data Logger mode (profile operations selected)
![]()
Change ac-9 Plus from Data logger to
ac-9 mode