WETStar
The meter has a unique flow tube design that lends itself to both pump-through and flow-through operation. It's power input of 7–15 VDC and 0–5 VDC analog output allow it to easily mate with existing CTD packages.
These miniature, low cost, low power optical instruments provide comparable performance to other fluorometers at a fraction of their cost, power requirements, and size. WETStar employs a novel optical flow tube design that lends itself to both pumpthrough and flow-through operation. It is easily mated with existing CTD packages and available with digital output.
Chlorophyll-a
Colored Dissolved Organic Matter
Uranine (fluorescein) & Rhodamine
Phycoerythrin
MECHANICAL
| Diameter | 6.9 cm |
| Length | 17.1 cm |
| Weight in air | 0.8 kg |
| Weight in water | 0.1 kg |
ELECTRICAL
| Input voltage | 7–15 VDC |
| Output, digital | 0–4095 counts |
| Output, analog | 0–5 V |
| Current draw, digital | 80 mA |
| Current draw, analog | 40 mA |
| Response time, digital | 0.125 sec |
| Response time, analog | 0.17 sec |
| Connector | MCBH6MP |
ENVIRONMENTAL
| Temperature range | 0 to 30 deg C |
| Depth rating | 600 m |
OPTICAL
Chlorophyll-a | ex/em: 460/695 nm |
| Sensitivity | 0.03 µg/I |
| Range | 0.03-75 µg/L |
CDOM | ex/em: 370/460 nm |
| Sensitivity | 0.100 ppb QSD |
| Ranges | 0-100, 0-250 ppb |
Uranine | ex/em: 485/530 nm |
| Sensitivity | 1 µg/I |
| Range | 0-4000 µg/I |
Rhodamine | ex/em: 470/590 nm |
Phycoerythrin | ex/em: 525/575 nm |
| Linearity (all) | 99% R2 |
Specifications subject to change without notice.
What should I use to lubricate the contacts on bulkhead connectors?
WET Labs recommends keeping the connectors on your meter lubricated using a good quality silicone spray lubricant, such as 3M Silicone Spray or Loctite 8021 spray. These can be purchased at your local hardware store. Avoid silicone grease. It is messy and attracts contaminants.
Do not use WD-40. It will destroy connectors.
What should I use to lubricate bulkhead connectors?
WET Labs recommends keeping the connectors on your meter lubricated using a good quality silicone spray lubricant, such as 3M Silicone Spray or Loctite 8021 spray. These can be purchased at your local hardware store. Avoid silicone grease. It is messy and attracts contaminants.
Do not use WD-40. It will destroy connectors.
Do you want WET Labs to review your data?
In order for data questions to be answered in an efficient manner, WET Labs requests:
- The serial number of the instrument data in question. The most current calibration information
can be verified and the instrument settings upon delivery can be checked.
- Description of the deployment configuration. Questions to consider:
- What are the current instrument settings?
- What is the instrument orientation?
- Are there possible obstructions?
- Is it being used in a mooring, a profiler, or being towed?
- What were the deployment date and geographical location?
- How far into the deployment was the problem seen?
- What power supply and cabling are being used?
- What was the instrument condition during the last field service or retrieval?
- Do you have any pictures?
- Raw data files of the physical data in question
- examples of good raw data.
- examples of questionable raw data .
- Information about how the raw data was processed.